Carbon-13


Method Summary

Isotope-ratio analysis for carbon-13 in solid-phase samples such as plant tissue starts with transformation to gas phase by extremely rapid and complete flash combustion of the sample material. Ionized combustion product (carbon dioxide) is mass-analyzed by means of differing mass/charge ratios among the various isotopic species of CO2. More detailsĀ  are available in the Process Overview.

Sample prep considerations

Plant samples must be oven-dried, ball-milled to less than 250 um particle size and weighed (~3 mg, with ug digits significant) into 5 x 5 mm tin capsules before combustion. Poor precision can often be traced to inadequate grinding that leaves fibrous matter or visible granules in the sample. Sample preparation for 13C analysis is detailed here.

Clients may send us unprocessed samples (e.g. leaves, stems, roots) or may provide any of the above processing themselves at a major reduction in price. Please refer to the Fee Schedule page.